Skip to main content

Cryotome (Thermo Fisher Shandon Cryotome FE/FSE)

Cryotome (Thermo Fisher Shandon Cryotome FE/FSE)

Introduction:

Precision cryostat for rapid sectioning of quick-frozen tissue specimens within a temperature-controlled chamber. Designed for routine cryo-microtomy in pathology and histology workflows.

 

Applications:

  • Routine cryosectioning of frozen tissue for diagnostic and research microscopy.

  • Sectioning using disposable or solid knives with anti-roll plate support.

 

Specifications

  • Function principle: Specimens are mounted to the specimen head using OCT and freezing. When mounted, the microtome within the chamber can be used for cryosectioning the frozen sample. This process requires training and some skill for obtaining high quality sections.

  • Controls & Interface:

    • Touch-screen display with micron selector, temperature readouts, specimen travel indicator
    • Advance/retract and rotary controls; standby mode
    • Heated window and cryobar; cryocassette clamp; waste bottle

 

Refrigeration:

  • The chamber and specimen head can cool down to aprx. -35C. Depending on the temperature of the room, etc.

  • Refrigerant: Isceon® 89

Safety:

  • NB. Sharp blade! The flywheel must be locked when mounting and adjusting the blade and sample.

  • Follow biohazard and chemical safety procedures.

Manual:

Last Updated 22.10.2025